Published on: January 20, 2022
Application of zero-phase digital filtering for effective denoising of field asymmetric waveform ion mobility
Junhui Li1,2, Wenqing Gao1,3, Huanming Wu1,2
1Zhejiang Provincial Key Laboratory of Advanced Mass Spectrometry and Molecular Analysis, Institute of Mass Spectrometry, Ningbo University, Ningbo, P. R. China.
Rationale:
Field asymmetric waveform ion mobility spectrometry (FAIMS) has a great potential to become a portable technology for rapid detection of chemical and biological agents. However, the ion current signals, measured at the exit of the planar FAIMS directly, may contain different types of noises. The peak information in the FAIMS spectrum, such as the compensation voltage (CV) value at the maximum peak intensity (CVP ) and the peak width at half maximum (Wh ), could not be accurately determined under the weak signal condition, which significantly limits the achievable instrument sensitivity, and there are no existing solutions to the problem.
Methods:
This study analyzed the noise type of FAIMS signal in detail, and three different signal processing algorithms, such as median filtering (MF), discrete wavelet transform (DWT), and zero-phase digital filtering (ZDF), were evaluated for their performance in denoising the FAIMS signal.
Results:
The results show that the standard deviation of CVp obtained from the signal denoised using ZDF algorithm is at least 31.82% smaller as compared to using MF and DWT algorithms. The standard deviation of Wh is at least 45.45% smaller using ZDF algorithm. Moreover, only ZDF algorithm can keep the percentage error for the CV value of the denoised signal to be within 0.50 ± 0.47% of the true CV value, implying the effectiveness of ZDF algorithm in denoising while retaining the integrity of the signal.
Conclusions:
The ZDF algorithm greatly reduces the analyte peak extraction error and improves the limit of detection in FAIMS measurements.
Related Concept Videos
Mass Analyzers: Common Types
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
Reconstruction of Signal using Interpolation
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences