Related Experiment Video
Updated: Oct 16, 2025

Radiolabeling and Quantification of Cellular Levels of Phosphoinositides by High Performance Liquid Chromatography-coupled Flow Scintillation
Published on: January 6, 2016
A CFD-based approach to optimize operating parameters of a flow-through scintillation cell for measurement of 220Rn
Tarun Kumar Agarwal1,2, Jitendra Jalindar Gaware3, Balvinder Kaur Sapra4,3
1Homi Bhabha National Institute, Anushaktinagar, Mumbai, 400094, India. tarun@barc.gov.in.
Abstract:
The measurements and monitoring of 222Rn/220Rn have been of emerging interest in occupational environments particularly in radium/thorium handling facilities and environments with monazite deposits for the inhalation dosimetry. The performance of a flow-through Lucas scintillation cell (LSC) for long run 220Rn measurements, depends upon the exact distribution pattern of 220Rn and its decay products in the LSC which can vary with the design of inlet path and flow rates. In this work, the CFD technique has been used to study the concentration profiles of 220Rn and its decay products in LSC for varying flow rates and inlet needle lengths. The variation of alpha production efficiency (ηα) is computed and analyzed for each case; aiming to select the best operating range of parameters for the optimum performance of LSC for 220Rn measurements. It is seen that LSC can be operated in the flow rate ranging from 0.6 to 1 lpm with inlet needle length varying from 22.5 to 45 mm for improved sensitivity.

