Electrostatic Discharge (ESD) as a Cause of Facial Nerve Stimulation After Cochlear Implantation: A Case Report
Smruti Rath1, Mica Glaun2, Claudia Emery3
1Baylor College of Medicine, Houston, TX, USA.
Objective:
To discuss persistent facial nerve stimulation (FNS) related to repeated electrostatic discharge (ESD) shock following cochlear implantation.
Methods:
Single case report with literature review.
Results:
FNS is a feared complication after cochlear implantation, occurring in approximately 7% of cases, with most patients having anatomic abnormalities. The presented case has no anatomical abnormalities but reported frequent environmental static shock. FNS during the first 1 to 3 seconds of processor attachment caused a significant decrease in the patient's quality of life, requiring subsequent re-implantation with full resolution.
Conclusions:
FNS is a complication of cochlear implantation that can cause a great deal of distress and discomfort. Frequent electrostatic discharge (ESD) contributed to device malfunctioning and FNS in a patient with otherwise normal anatomy and should be avoided if possible.