Related Experiment Video
Updated: Oct 16, 2025

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Application of a six sigma model to evaluate the analytical performance of urinary biochemical analytes and design a
Qian Liu1, Guangrong Bian1, Xinkuan Chen2
1Department of Laboratory Medicine, The Second People's Hospital of Lianyungang, Lianyungang, China.
Background:
The six sigma model has been widely used in clinical laboratory quality management. In this study, we first applied the six sigma model to (a) evaluate the analytical performance of urinary biochemical analytes across five laboratories, (b) design risk-based statistical quality control (SQC) strategies, and (c) formulate improvement measures for each of the analytes when needed.
Methods:
Internal quality control (IQC) and external quality assessment (EQA) data for urinary biochemical analytes were collected from five laboratories, and the sigma value of each analyte was calculated based on coefficients of variation, bias, and total allowable error (TEa). Normalized sigma method decision charts for these urinary biochemical analytes were then generated. Risk-based SQC strategies and improvement measures were formulated for each laboratory according to the flowchart of Westgard sigma rules, including run sizes and the quality goal index (QGI).
Results:
Sigma values of urinary biochemical analytes were significantly different at different quality control levels. Although identical detection platforms with matching reagents were used, differences in these analytes were also observed between laboratories. Risk-based SQC strategies for urinary biochemical analytes were formulated based on the flowchart of Westgard sigma rules, including run size and analytical performance. Appropriate improvement measures were implemented for urinary biochemical analytes with analytical performance lower than six sigma according to the QGI calculation.
Conclusions:
In multilocation laboratory systems, a six sigma model is an excellent quality management tool and can quantitatively evaluate analytical performance and guide risk-based SQC strategy development and improvement measure implementation.
Related Concept Videos
Urine Studies I: Urinalysis
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Data Validation
Key parameters for method validation include:
Introduction to Statistical Process Control
Urine Studies II: Urine Culture and Sensitivity Test
Bioequivalence Data: Statistical Interpretation