Real-world outcomes of two different sensor-augmented insulin pumps with predictive low glucose suspend function in
J Moreno-Fernandez1, P Beato-Vibora2, P Olvera3
1Endocrinology and Nutrition Service, Ciudad Real General University Hospital, Ciudad Real, Spain.
Aim:
To analyse the real-life outcomes of two sensor-augmented pumps (SAP) with predictive low glucose suspend (PLGS) function, Medtronic Minimed 640G™ with SmartGuard (MM640G) and Tandem T Slim X2™ with Basal-IQ™ (TTSX2), in Type 1 Diabetes Mellitus (T1DM) patients.
Methods:
Observational cross-sectional study using data obtained from computerized clinical records. All T1DM patients on TTSX2 therapy were compared (1:1) with MM640G treated patients selected through stratified sampling. Primary efficacy outcome was to describe time in rage (TIR, 70-180 mg/dL, 3.9-10 mmol/L) interstitial glucose differences according to a non-inferiority hypothesis with TTSX2 compared to MM640G.
Results:
Forty-four patients were analyzed (female 66%). Mean age was 38.9 yrs. (range 23-59 yrs.) and mean diabetes duration was 23.4 ± 9.2 yrs. Patients treated with TTSX2 showed a numerically slightly lower, but non-statistically significantly different, TIR from the MM640G pump group (64.9 ± 16.4% vs. 72.4 ± 17.0%, P = 0.108). Similarly, we did no find differences in HbA1c between T1D patients treated with TTSX2 and MM640G (6.8 ± 1.0% vs. 7.0 ± 0.9%, 51 ± 11 mmol/mol vs. 53 ± 10 mmol/mol, P = 0.312). Moreover, rest of evaluated glycemic outcomes were similar between both treatment groups.
Conclusions:
Patients using two different SAP with PLGS automatic function showed similar glycaemic control in a real-world scenario. NCT04741685.
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Diabetes Mellitus: Type 2 and Gestational
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Hypoglycemia and Glucagon
Glucagon-like Receptor Agonists
GLP-1, when administered in high doses intravenously, triggers insulin secretion, inhibits glucagon release, slows gastric emptying, reduces food intake, and restores normal insulin secretion. However, its rapid inactivation by...