A comparison of clinical outcomes and optical performance between monofocal and new monofocal with enhanced
Jungah Huh1,2, Youngsub Eom1,3, Seul Ki Yang4,5
1Department of Ophthalmology, Korea University College of Medicine, Seoul, Republic of Korea.
Background:
To compare clinical outcomes and optical performance of a new monofocal with enhanced intermediate function intraocular lenses (IOLs) with that of conventional monofocal IOLs.
Methods:
Sixty eyes of 30 patients who underwent phacoemulsification with bilateral implantation of the ICB00 (15 patients) or ZCB00 (15 patients) IOLs were enrolled. Binocular corrected distance visual acuity (CDVA), distance corrected near visual acuity (DCNVA), and distance corrected intermediate visual acuity (DCIVA) were measured at 4 weeks after surgery. Patient satisfaction for overall, near, intermediate, and distance vision were assessed. The binocular defocus curves were measured. The root mean square of modulation transfer function (MTFRMS) was measured in the optical bench study.
Results:
The mean binocular DCIVA was significantly better in the ICB00 group (0.01 logMAR) compared to the ZCB00 group (0.13 logMAR), but CDVA and DCNVA were not. The patient satisfaction for near and intermediate vision was significantly higher in the ICB00 group compared to the ZCB00. However, there was no difference in patient satisfaction for overall and distance vision between two groups. The defocus curves showed that mean visual acuity of the ICB00 group was significantly better than that of the ZCB00 group at between - 1.00 D to - 3.00 D of defocus. The ICB00 IOL had higher MTFRMS values at between - 0.50 D to - 2.00 D of defocus compared to the ZCB00 IOL.
Conclusions:
The ICB00 IOL provides better binocular intermediate vision and higher satisfaction for near and intermediate vision than the ZCB00 IOL while maintaining excellent distance vision.
More Related Videos
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
06:19Comparison of Three Clinical Stereoscopic Methods for Measuring Binocular Visual Function During Amblyopic Treatment in Unilateral Amblyopia
Published on: September 27, 2024
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment
Focusing of Light in the Eye
Glaucoma: Overview