Related Experiment Video
Updated: Oct 16, 2025

Hyperpolarized 13C Metabolic Magnetic Resonance Spectroscopy and Imaging
Published on: December 30, 2016
Optimization of spectrally selective 180° radiofrequency pulse timings in J-difference editing (MEGA) of lactate
Sandeep K Ganji1,2,3, Zhongxu An1, Vivek Tiwari1
1Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
Purpose:
J-Difference editing (MEGA) provides an effective spectroscopic means of selectively measuring low-concentration metabolites having weakly coupled spins. The fractional inphase and antiphase coherences are determined by the radiofrequency (RF) pulses and inter-RF pulse intervals of the sequence. We examined the timings of the spectrally selective editing 180° pulses (E180) in MEGA-PRESS to maximize the edited signal amplitude in lactate at 3T.
Methods:
The time evolution of the lactate spin coherences was analytically and numerically calculated for non-volume localized and single-voxel localized MEGA sequences. Single-voxel localized MEGA-PRESS simulations and phantom experiments were conducted for echo time (TE) 60-160 ms and for all possible integer-millisecond timings of the E180 pulses. Optimized E180 timings of 144, 103, and 109 ms TEs, tailored with simulation and phantom data, were tested in brain tumor patients in vivo. Lactate signals, broadened to singlet linewidths (~6 Hz), were compared between simulation, phantom, and in vivo data.
Results:
Theoretical and experimental data indicated consistently that the MEGA-edited signal amplitude and width are sensitive to the E180 timings. In volume-localized MEGA, the lactate peak amplitudes in E180-on and difference spectra were maximized at specific E180 timings for individual TEs, largely due to the chemical-shift displacement effects. The E180 timings for maximum lactate peak amplitude were different from those of maximum inphase coherence in in vivo linewidth situations.
Conclusion:
In in vivo MEGA editing, the E180 pulse timings can be effectively used for manipulating the inphase and antiphase coherences and increasing the edited signal amplitude, following TE optimization.
More Related Videos
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
10:54Dissolution Dynamic Nuclear Polarization Instrumentation for Real-time Enzymatic Reaction Rate Measurements by NMR
Published on: February 23, 2016
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)