Background:
Balloon-expandable pulmonary valves are usually not suitable for dilated native outflow tracts.
Methods:
Indian Venus P-valve registry was retrospectively analyzed for efficacy, complications, and midterm outcomes. Straight valve was used in prestented conduits in patients with right ventricular pressure above two-thirds systemic pressure and/or right ventricular dysfunction. Flared valve 1-4 mm larger than balloon waist was used in native outflow in symptomatic patients, large ventricular volumes, and ventricular dysfunction.
Objectives:
A self-expanding porcine pericardial Venus P-valve is available in straight and flared designs..
Results:
Twenty-nine patients were included. Straight valve was successful in all seven conduits, reducing gradients significantly, including one patient with left pulmonary artery (LPA) stent. Flared valve was successfully implanted in 20 out of 22 native outflow tracts. Sharp edges of the older design contributed to two failures. Complications included two migrations with one needing surgery, endocarditis in one, insignificant wire-frame fractures in three, and groin vascular complication in one patient. There were no deaths or valve-related reinterventions at a mean follow-up of 47.8 ± 24.5 months (1-85 months). Modifications of technique succeeded in three patients with narrow LPA. There was significant improvement in symptoms, right ventricular volume, and pulmonary regurgitant fraction.
Conclusion:
Straight and flared Venus P-valves are safe and effective in appropriate outflow tracts. Straight valve is an alternative to balloon-expandable valves in stenosed conduits. Flared valve is suitable for large outflows up to 34 mm, including patients with LPA stenosis. Recent design modifications may correct previous technical failures. Studies should focus on durability and late complications.
More Related Videos
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Stenosis I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Venous Return
What is Venous Return?
Venous return refers to the rate at which blood flows back to the heart from the body's peripheral veins. It's an integral part of the circulatory system...