Bismuth Pelvic X-Ray Shielding Reduces Radiation Dose Exposure in Pediatric Radiography
Bow Wang1, Chien-Yi Ting2, Cheng-Shih Lai1,2,3
1Department of Diagnostic Radiology, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan 701, Taiwan.
Background:
Radiation using conventional X-ray is associated with exposure of radiosensitive organs and typically requires the use of protection. This study is aimed at evaluating the use of bismuth shielding for radiation protection in pediatric pelvic radiography. The effects of the anteroposterior and lateral bismuth shielding were verified by direct measurements at the anatomical position of the gonads.
Methods:
Radiation doses were measured using optically stimulated luminescence dosimeters (OSLD) and CIRS ATOM Dosimetry Verification Phantoms. Gonad radiographs were acquired using different shields of varying material (lead, bismuth) and thickness and were compared with radiographs obtained without shielding to examine the effects on image quality and optimal reduction of radiation dose. All images were evaluated separately by three pediatric orthopedic practitioners.
Results:
Results showed that conventional lead gonadal shielding reduces radiation doses by 67.45%, whereas dose reduction using one layer of bismuth shielding is 76.38%. The use of two layers of bismuth shielding reduces the dose by 84.01%. Using three and four layers of bismuth shielding reduces dose by 97.33% and 99.34%, respectively. Progressively lower radiation doses can be achieved by increasing the number of bismuth layers. Images obtained using both one and two layers of bismuth shielding provided adequate diagnostic information, but those obtained using three or four layers of bismuth shielding were inadequate for diagnosis.
Conclusions:
Bismuth shielding reduces radiation dose exposure providing appropriate protection for children undergoing pelvic radiography. The bismuth shielding material is lighter than lead, making pediatric patients more comfortable and less apt to move, thereby avoiding repeat radiography.
Related Concept Videos
Pharmacokinetics in Pediatric Patients: Overview and Drug Absorption
Pharmacokinetics in Pediatric Patients: Drug Excretion
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
X-ray Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...