A heparin-functionalized covered stent prepared by plasma technology
Related Experiment Video
Updated: Oct 16, 2025

Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents
Published on: October 26, 2016
Qing Zhu1, Ping Ye1, Fang Guo1
147863University of Shanghai for Science and Technology, Shanghai, China.
Abstract:
In this study, the surface of the covered stent was treated by plasma technology to introduce amino functional groups, and glutaraldehyde and heparin were successfully grafted to prepare a heparin-functionalized covered stent (HPLCS). The preparation parameters such as plasma treatment power, plasma treatment time, concentration of glutaraldehyde and heparin, and pH of heparin solution were studied in detail. The functionalized heparin covered stent can make the titer of heparin reach 1.23 ± 0.03 IU/cm2. In animal experiments, after implantation in pigs for 6 months, the titer of heparin can still reach 0.93 ± 0.05 IU/cm2. This work provides a good method for preparing heparin covered stent.