U C Onubogu1, W Wonodi1, B A West1
1Department of Paediatrics and Child Health, Rivers State University Teaching Hospital, Port Harcourt, Rivers State. Nigeria.
Background:
Body temperature is an important vital sign in clinical practice which can be measured via electronic contact thermometers and infrared non-contact thermometers.
Objective:
To compare temperature readings taken by non contact infrared thermometer with the conventional digital axillary, rectal and oral temperature readings as well as the influence of environmental temperature on noncontact infrared thermometer readings.
Methodology:
A prospective study carried out in the Paediatric outpatient clinic of the Rivers State University Teaching Hospital, Nigeria from September 2020 to December 2020. Infrared noncontact forehead and jugular temperatures along with contact axillary, oral and rectal temperatures at a recorded atmospheric temperature and pressure were measured. Data collected was analysed.
Results:
A total of 247 children aged 1month to 16 years were enrolled, the mean differences of the temperature pairs of contact and non-contact thermometry ranged from 0.45 - 0.77°C (1.64, -1.81°C) 95% LoA. The highest mean difference was found between infrared forehead and axillary [MD; 0.45(1.64,-0.73°C) 95%LoA] temperatures. There was a significant positive correlation between the mean difference of infrared forehead/ rectal temperature and atmospheric temperature (r = 0 .211 p = 0.029). Linear regression model showed that infrared forehead temperature of 37.1°C was equivalent to rectal temperature of 38°C and axillary of 37.4°C which is the standard cut off for fever. Infrared jugular of 37.2°C was equivalent to rectal of 38°C and axillary temperature of 37.4°C was equivalent to infrared jugular of 37.03°C all at a mean atmospheric temperature of 28.3±1.8°C.
Conclusion:
The mean difference by which infrared noncontact thermometry predicts core temperatures may differ based on atmospheric temperature. Infrared non-contact forehead thermometer reading of 37.1°C could be considered as the fever cut off for non-contact forehead thermometry in Nigeria in regions where the mean atmospheric temperature is 28.3±1.8°C.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4: