Related Experiment Video
Updated: Oct 16, 2025

Preparation and High-temperature Anti-adhesion Behavior of a Slippery Surface on Stainless Steel
Published on: March 29, 2018
The packaging and clean method contribute to insulation failure of electrosurgical instruments
Ying Zhang1, Yanyan Zhang1, Yafei Wang2
1Central Sterile Supply Department, Cangzhou Central Hospital, Cangzhou, Hebei, China.
Abstract:
With the rapid development of medical technology, the use of electrosurgical instruments is dramatically increased in various types of surgery. However, the damage of the insulation layer of the reusable electrosurgical instrument often causes surgical accidents. The procedures of packaging and cleaning contribute to many damages to insulating layer of reusable electrosurgical instruments.Various types of reusable electrosurgical instruments were detected for insulation failures, conduction failures, short-circuit by using a high-voltage detector, DIATEG (Morgate company). In addition, reusable electrosurgical instruments were detected for insulation failures after packaging and cleaning by different procedures.13.1% (129/740) electrosurgical instruments had an insulation test failure; 6.2% (9/146) monopolar wires were with conduction failure; and 7.7% (16/207) bipolar wires were with short-circuit. Different packaging and cleaning procedures contribute to various degrees of damages to insulating property of reusable electrosurgical instruments.Insulation failure was a wide problem of reusable electrosurgical instruments, while fixed packaging method and mild cleaning procedures result in fewer damages to insulating property of reusable electrosurgical instruments.
More Related Videos
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Insulation Coordination
Hand hygiene
Hand washing...