ALADA Dose Optimization in the Computed Tomography of the Temporal Bone: The Diagnostic Potential of Different
Barbara Kofler1, Laura Jenetten2, Annette Runge1
1Department of Otorhinolaryngology, Medical University of Innsbruck, Anichstrasse 35, 6020 Innsbruck, Austria.
Objective:
Repeated computed tomography (CT) is essential for diagnosis, surgical planning and follow-up in patients with middle and inner ear pathology. Dose reduction to "as low as diagnostically acceptable" (ALADA) is preferable but challenging. We aimed to compare the diagnostic quality of images of subtle temporal bone structures produced with low doses (LD) and reference protocols (RP).
Methods:
Two formalin-fixed human cadaver heads were scanned using a 64-slice CT scanner and cone-beam CT (CBCT). The protocols were: RP (120 kV, 250 mA, CTDIvol 83.72 mGy), LD1 (100 kV, 80 mA, CTDIvol 26.79 mGy), LD2 (100 kV, 35 mA, CTDIvol 7.66 mGy), LD3 (80 kV, 40 mA, CTDIvol 4.82 mGy), and CBCT standard protocol. Temporal bone structures were assessed using a 5-point scale.
Results:
A median score of ≥2 was achieved with protocols such as the tendons of m. tensor tympani (RP/LD1/LD2/CBCT) and m. stapedius (CBCT), the incudostapedial joint (RP/LD1/CBCT), the incudomalleolar joint (RP/LD1/LD2/CBCT), the stapes feet (RP/LD1/CBCT), the stapes head (RP/LD1/LD2/CBCT), the tympanic membrane (RP/LD1/LD2/CBCT), the lamina spiralis ossea (none), the chorda tympani (RP/LD1/CBCT), and the modiolus (RP/LD1/LD2/CBCT). Adaptive statistical iterative reconstructions did not show advantages over the filtered back projection.
Conclusions:
LD protocols using a CTDIvol of 7.66 mGy may be sufficient for the identification of temporal bone structures.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...