Visual Evaluation of Image Quality of a Low Dose 2D/3D Slot Scanner Imaging System Compared to Two Conventional
Ahmed Jibril Abdi1,2, Bo Mussmann1,3, Alistair Mackenzie4
1Department of Clinical Research, University of Southern Denmark, 5000 Odense, Denmark.
Abstract:
The purpose of this study was to assess the image quality of the low dose 2D/3D slot scanner (LDSS) imaging system compared to conventional digital radiography (DR) imaging systems. Visual image quality was assessed using the visual grading analysis (VGA) method. This method is a subjective approach that uses a human observer to evaluate and optimise radiographic images for different imaging technologies.
Methods And Materials:
ten posterior-anterior (PA) and ten lateral (LAT) images of a chest anthropomorphic phantoms and a knee phantom were acquired by an LDSS imaging system and two conventional DR imaging systems. The images were shown in random order to three (chest) radiologists and three experienced (knee) radiographers, who scored the images against a number of criteria. Inter- and intraobserver agreement was assessed using Fleiss' kappa and weighted kappa.
Results:
the statistical comparison of the agreement between the observers showed good interobserver agreement, with Fleiss' kappa coefficients of 0.27-0.63 and 0.23-0.45 for the chest and knee protocols, respectively. Comparison of intraobserver agreement also showed good agreement with weighted kappa coefficients of 0.27-0.63 and 0.23-0.45 for the chest and knee protocols, respectively. The LDSS imaging system achieved significantly higher VGA image quality compared to the DR imaging systems in the AP and LAT chest protocols (p < 0.001). However, the LDSS imaging system achieved lower image quality than one DR system (p ≤ 0.016) and equivalent image quality to the other DR systems (p ≤ 0.27) in the knee protocol. The LDSS imaging system achieved effective dose savings of 33-52% for the chest protocol and 30-35% for the knee protocol compared with DR systems.
Conclusions:
this work has shown that the LDSS imaging system has the potential to acquire chest and knee images at diagnostic quality and at a lower effective dose than DR systems.
More Related Videos
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography