Optimizing Parameters for Smoke Evacuation
Jusleen Ahluwalia1, Emily de Golian, Joyce Yuan
1All authors are affiliated with the Department of Dermatology, University of California, San Diego School of Medicine, La Jolla, California.
Background:
Current literature lacks recommendations regarding the ideal organization of the smoke evacuation system to minimize inhalation of surgical smoke.
Objective:
This study determines optimal parameters of the smoke evacuation system with respect to the surgical field.
Materials And Methods:
This study was conducted in an outpatient surgical facility at a tertiary care center. After 30 seconds of continuous electrocautery of tissue, particulate measurements were recorded using the TSI DustTrak Aerosol Monitor 8520 (>2.5 μm particles) and the TSI P-Trak Ultrafine Particle Counter 8525 (<1 μm particles) while changing the angle and distance of the smoke evacuation system.
Results:
Particulate matter measurements were lower when suction angle was at 45° than at 90°. For both small-sized and large-sized particles, the lowest particulate matter was recorded when the evacuator was maintained at a 45° angle, 2 to 4 inches away from the cauterizing procedure. Particulate matters dramatically increased after increasing the distance of the smoke evacuator beyond 8 inches from the procedural site.
Conclusion:
In an effort to reduce smoke inhalation, the authors recommend that smoke evacuation should be placed at a 45° suction angle, no further than 8 inches away from the surgical site, while preserving the surgeon's field of vision.