Related Experiment Video
Updated: Oct 15, 2025

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
The Application of Electrical Impedance Tomography During the Ventilator Weaning Process
Guan Wang1, Lei Zhang1, Bin Li1
1Department of Critical Care, Beijing Luhe Hospital, Capital Medical University, Beijing, People's Republic of China.
Background:
This study proposes the investigation of electrical impedance tomography (EIT) as a useful predictor for ventilator weaning.
Methods:
The study design was a nested case-control study and patients who were admitted to the intensive care unit and underwent their first tracheal intubation were enrolled. Those who successfully completed ventilator weaning and extubation after the first spontaneous breathing trial (SBT) were included in the weaning success group, while those who did not pass the SBT or received secondary intubation within 48 hours were included in the weaning failure group. In both groups, EIT was adopted to record the monitoring data in three phases: before the SBT (pre-SBT), during the SBT (SBT), and after the SBT (post-SBT).
Results:
A total of 53 patients were enrolled, including 41 cases in the weaning success group and 12 cases in the weaning failure group. The logistic regression analysis showed that the pre-SBT global impedance (GI) and the SBT region of interest 2 (ROI2) were significantly higher in the weaning success group than in the weaning failure group (p = 0.0001 and p = 0.002). The pre-SBT GI predicted weaning success with a sensitivity of 0.524, a specificity of 0.818, a p-value of 0.0496, and a 95% confidence interval (CI) of 0.001-0.978. The sensitivity, specificity, p-value, and 95% CI for the SBT ROI2 were 1, 0.595, 0.0164, and 1.010-1.108, respectively.
Conclusion:
For patients without contraindications to EIT, the application of EIT is recommended to be added to the existing evaluation system for ventilator weaning, as it could help improve the weaning success rate. Further cohort studies are needed to investigate the actual efficacy of EIT after it has been added to the evaluation system.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....