Equipment failure during mandibular staple bone plate insertion and a technique to resolve it
Summary
A fractured set screw on a mandibular staple bone plate drill guide prevented apparatus removal. Cutting director rods successfully resolved this surgical complication.
Area of Science:
- Oral and Maxillofacial Surgery
- Biomedical Engineering
- Orthopedic Surgery
Background:
- The mandibular staple bone plate is utilized in specific mandibular fracture repairs.
- Drill guides are essential instruments for accurate placement of bone plates.
- Set screws are critical components for securing surgical instruments during procedures.
Observation:
- Intraoperative fracture of the set screw within the drill guide was observed.
- This fracture led to the surgical apparatus being irremovable.
- The mandibular staple bone plate system was implicated in the complication.
Findings:
- The fractured set screw obstructed the removal of the drill guide.
- A novel problem-solving approach was required to address the intraoperative complication.
- Successful removal of the apparatus was achieved by sectioning the director rods.
Implications:
- This case highlights a potential failure mode in surgical drill guide design.
- Clinicians should be prepared for unexpected instrument failures during surgery.
- Alternative removal techniques may be necessary when standard methods are impeded.