Simultaneous 3D-TOF angiography and 4D-flow MRI with enhanced flow signal using multiple overlapping thin slab
Dahan Kim1,2, Laura Eisenmenger3, Patrick Turski3
1Department of Physics, University of Wisconsin, Madison, Wisconsin, USA.
Purpose:
To investigate the fusion of 3D time-of-flight principles into 4D-flow MRI to enhance vessel contrast and signal without an exogenous contrast agent, enabling simultaneous in-flow based angiograms.
Methods:
A 4D-flow MRI technique was developed consisting of multiple overlapping slabs with intermittent magnetization transfer preparation. The scan time penalty associated with multiple slab acquisitions was mitigated by using undersampled distributed spiral trajectories and compressed sensing reconstruction. A flow phantom was used to characterize in-flow enhancement, velocity noise improvement, and flow rate measurements against the single-slab 4D-flow MRI. In a patient-volunteer cohort (n = 15), magnitude-based angiograms were radiologically evaluated against 3D time-of-flight, and velocity measurements were compared pixel-wise against single-slab and contrast-enhanced 4D-flow MRI.
Results:
Multiple-slab acquisitions, together with magnetization transfer preparation, substantially improved vessel signal, contrast, and vessel conspicuity in magnitude angiograms. Both clinical 3D time-of-flight and the proposed technique produced equivalent vessel depictions with no statistically significant difference (p < .1). Both techniques also produced clear depictions of brain aneurysms in all patients; however, very small vessels tended to show reduced conspicuity in the proposed technique. Velocity measurements agreed with contrast-enhanced and single-slab scans with high correlations (R2 = 0.941-0.974) and agreements (slopes = 0.994-1.071). Slab boundary and magnetization transfer-related artifacts were not observed in velocity measurements, and velocity noise was reduced with in-flow enhancement over single-slab scans (phantom).
Conclusion:
The vessel signal and contrast can be improved in 4D-flow MRI without exogenous contrast agents by utilizing in-flow enhancement, efficient sampling, and compressed sensing. The in-flow enhancement also enables simultaneous 3D time-of-flight angiograms useful for flow quantification and diagnosis.