Related Experiment Video
Updated: Oct 15, 2025

Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
Correlations between home sleep apnea tests and polysomnography outcomes do not fully reflect the diagnostic accuracy
Frederik Massie1, Bart Van Pee1, Jeroen Bergmann1
1Natural Interaction Lab, Thom Building, Department of Engineering, University of Oxford, Oxford, United Kingdom.
Study Objectives:
The clinical performance of home sleep apnea tests (HSATs) can be described by their (diagnostic) accuracy, defined as the percentage agreement with the obstructive sleep apnea severity category (normal, mild, moderate, and severe) based on polysomnography. Rather than reporting on accuracy, there has been a strong reliance in the literature to report correlation coefficients between the apnea-hypopnea index of HSATs and polysomnography to support claims of diagnostic performance. This is surprising, as it has been well described that correlation coefficients are inadequate to evaluate equivalence between 2 parameters. The aim of this study was to systematically investigate the magnitude of the discrepancies between correlation coefficients and diagnostic accuracy reported in or retrievable from HSAT validation studies.
Methods:
We compared the discrepancy between accuracy and apnea-hypopnea index correlation coefficients of all validation papers that met the inclusion criteria. A total of 20 papers were retained, representing a participant pool of 1,652.
Results:
The weighted average apnea-hypopnea index correlation across all 20 papers was 0.82 and the weighted average accuracy was 0.61, highlighting a discrepancy of 0.21 and an overall misdiagnosis rate of 39%.
Conclusions:
The results of our study confirm the need for increased scientific rigor in selecting primary performance endpoints to support clinical performance claims of HSATs.
Citation:
Massie F, Van Pee B, Bergmann J. Correlations between home sleep apnea tests and polysomnography outcomes do not fully reflect the diagnostic accuracy of these tests. J Clin Sleep Med. 2022;18(3):871-876.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
07:54Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Related Concept Videos
Correlations
Sleep Apnea
The condition is more prevalent among...
Reliability and Validity
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5%...
Sensitivity, Specificity, and Predicted Value
Sensitivity is the...
Errors occurring during blood pressure monitoring
Several factors...