Rapid visualization and detection of Staphylococcus aureus based on loop-mediated isothermal amplification.
Chuan Wu1, Yuanyuan Zeng1, Yang He2
1Chongqing Key Laboratory of Sichuan-Chongqing Co-Construction for Diagnosis and Treatment of Infectious Diseases Integrated Traditional Chinese and Western Medicine, College of Medical Technology, Chengdu University of Traditional Chinese Medicine, Chengdu, 611137, Sichuan, China.
A new loop-mediated isothermal amplification (LAMP) assay offers rapid and accurate detection of Staphylococcus aureus. This colorimetric method provides a sensitive tool for diagnosing S. aureus infections, outperforming conventional PCR.
Area of Science:
- Microbiology
- Molecular Biology
- Diagnostic Assays
Background:
- Staphylococcus aureus is a significant clinical pathogen causing various infections.
- Rapid and reliable detection assays for S. aureus are crucial for timely diagnosis and treatment.
- Existing diagnostic methods may lack the speed or sensitivity required for early detection.
Purpose of the Study:
- To develop a closed-tube, visual, colorimetric loop-mediated isothermal amplification (LAMP) assay for detecting Staphylococcus aureus.
- To optimize the LAMP reaction conditions for enhanced specificity and sensitivity.
- To evaluate the diagnostic accuracy of the developed LAMP assay compared to conventional methods using clinical isolates.
Main Methods:
- Development of a closed-tube loop-mediated isothermal amplification (LAMP) assay.
- Utilized hydroxy naphthol blue (HNB) as a colorimetric indicator for visual detection.
- Optimized reaction parameters including temperature, Mg2+, dNTP, HNB concentrations, and incubation time.
- Validated specificity against 14 non-S. aureus strains and determined the limit of detection (LOD).
- Evaluated diagnostic efficacy using 220 S. aureus and 149 non-S. aureus clinical isolates.
Main Results:
- The optimized LAMP assay demonstrated 100% specificity for S. aureus with no cross-reactivity.
- The limit of detection (LOD) was as low as 8 copies/reaction for plasmid DNA or 400 CFU/reaction for S. aureus.
- LAMP achieved a tenfold lower LOD compared to conventional PCR.
- The assay showed a high diagnostic accuracy of 99.46% on clinical isolates.
Conclusions:
- The developed colorimetric LAMP assay is a highly specific and sensitive method for detecting Staphylococcus aureus.
- This assay offers a reliable and rapid alternative for the diagnosis of S. aureus infections.
- LAMP presents a promising tool for clinical settings requiring swift and accurate pathogen identification.