Investigation of Eye Lens Dose Estimate based on AAPM Report 293 in Head Computed Tomography
Choirul Anam1, Winda Kusuma Dewi2, Masdi Masdi3
1PhD, Department of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Jl. Prof. Soedarto SH, Semarang 50275, Central Java, Indonesia.
Background:
Estimation of eye lens dose is important in head computed tomography (CT) examination since the eye lens is a sensitive organ to ionizing radiation.
Objective:
The purpose of this study is to compare estimations of eye lens dose in head CT examinations using local size-specific dose estimate (SSDE) based on size-conversion factors of the American Association of Physicists in Medicine (AAPM) Report No. 293 with those based on size-conversion factors of the AAPM Report No. 220.
Material And Methods:
This experimental study is conducted on a group of patients who had undergone nasopharyngeal CT examination. Due to the longitudinal (z-axis) dose fluctuation, the average global SSDE and average local SSDE (i.e. particular slices where the eyes are located) were investigated. All estimates were compared to the measurement results using thermo-luminescent dosimeters (TLDs). The estimated and measured doses were implemented for 14 patients undergoing nasopharyngeal CT examination.
Results:
It was found that the percentage differences of the volume CT dose index (CTDIvol), average global SSDE based on AAPM No. 220 (SSDEo,g), average local SSDE based on AAPM No. 220 (SSDEo,l), average global SSDE based on AAPM No. 293 (SSDEn,g) and average local SSDE based on AAPM No. 293 (SSDEn,l) against the measured TLD doses were 22.5, 21.7, 15.0, 9.3, and 2.1%, respectively. All comparisons between dose estimates and TLD measurements gave p-values less than 0.001, except for SSDEn,l (p-value = 0.566).
Conclusion:
SSDE based on AAPM Report No. 293 can be used to accurately estimate eye lens radiation doses by performing the calculations on a number of specific slices containing the eyes.
More Related Videos
09:11Assessment of Global Ocular Structure Following Spaceflight Using a Micro-Computed Tomography Micro-CT Imaging Method
Published on: October 27, 2020
10:23Author Spotlight: A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
Published on: June 23, 2023
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...