Improved Sensitivity of Ultrasound-Based Subharmonic Aided Pressure Estimation Using Monodisperse Microbubbles
Wim van Hoeve1, Miguel de Vargas Serrano1, Lisa Te Winkel1
1Solstice Pharmaceuticals B.V, Enschede, Netherlands.
Monodisperse microbubbles significantly enhance Subharmonic Aided Pressure Estimation (SHAPE) sensitivity for noninvasive hydrostatic pressure measurement. This advancement nearly doubles the sensitivity compared to current clinical methods.
Area of Science:
- Biomedical Engineering
- Acoustic Imaging
- Medical Device Technology
Background:
- Subharmonic Aided Pressure Estimation (SHAPE) is a valuable noninvasive technique for measuring hydrostatic fluid pressures.
- Current SHAPE methods utilize polydisperse microbubbles, which may limit sensitivity and accuracy.
- Improving SHAPE sensitivity is crucial for broader clinical applications and enhanced diagnostic capabilities.
Purpose of the Study:
- To investigate the potential of monodisperse microbubbles to improve SHAPE sensitivity.
- To characterize the acoustic properties and stability of monodisperse microbubble populations.
- To evaluate the performance of monodisperse microbubbles in a hydrostatic pressure-controlled environment.
Main Methods:
- Monodisperse microbubbles were fabricated using microfluidics and characterized for size distribution and stability.
- Acoustic attenuation and resonance frequencies were measured to determine microbubble properties.
- SHAPE sensitivity was assessed using a clinical ultrasound scanner in a controlled pressure bath.
Main Results:
- Successfully created monodisperse microbubble suspensions (2.4–5.3 μm) with stable size distributions over 24 hours.
- Resonance frequencies correlated strongly with microbubble diameter (R² > 0.99).
- Monodisperse bubbles exhibited nearly double the SHAPE sensitivity (-0.17 dB/mmHg) compared to commercial polydisperse bubbles.
Conclusions:
- Monodisperse microbubbles offer a significant improvement in SHAPE sensitivity.
- This advancement holds promise for more accurate noninvasive hydrostatic pressure measurements.
- Further development could lead to enhanced diagnostic tools in various clinical settings.