Related Experiment Video
Updated: Oct 14, 2025

Simulator Training for Endovascular Neurosurgery
Published on: May 6, 2020
Acute frame coil migration during filling coil retrieval in a cerebral aneurysm embolization case: A possible result
Omer Doron1, Eran Meirowitz1, Eitan Abergel1
1Rambam Health Care Campus, Rappaport Faculty of Medicine, the Technion - Israel Institute of Technology, Haifa, Israel.
Abstract:
Coil migration is a known complication of endovascular coiling of cerebral aneurysms. We report an acute coil migration occurring during coiling of an unruptured anterior choroidal aneurysm, while a separate coil was retrieved into the microcatheter concomitantly without direct contact between the coils. The "pulling" of a previously deployed stable coil is presented as an adverse effect that should be noted. This case exemplifies that not only direct entanglement or erroneous malposition can generate acute migration, but possibly, also alteration of physical forces during coil retrieval. A potential mechanism, similar to a venturi effect, caused by a quick suction within the microcatheter at its distal end, is suggested, and the clinical relevance of the case is discussed.
Related Concept Videos
Aneurysm IV: Nursing Management
Aneurysm I: Introduction
Aneurysm III: Interprofessional Care