Total lab automation: sample stability of clinical chemistry parameters in an automated storage and retrieval module
Kobe Vercruysse1, Stijn Lambrecht1, Matthijs Oyaert1
1Department of Laboratory Medicine, Ghent University Hospital, Ghent, Belgium.
Objectives:
Automated storage and retrieval modules (SRM), as part of total lab automation (TLA) systems, offer tremendous practical and economic benefits. In contrast to manual storage systems, SRMs indicate continuous motion of samples and may leave samples prone to temperature fluctuations. This study investigates analyte stability in serum and heparin plasma within an automated storage module.
Methods:
The stability of 28 common biochemistry analytes was investigated using 57 freshly obtained routine serum samples and 42 lithium-heparin plasma samples. Following baseline measurement, samples were stored at 2-8 °C in the automated SRM of the Accelerator a3600 TLA and reanalyzed at fixed time points (2, 4, 8, 12, 24, 48 and 72 h) on the Abbott Architect c16000 chemistry analyzer. The concentration at each time point was expressed as %-difference to the baseline value and mean results were compared to the criteria for desirable bias derived from the biological variation database.
Results:
Nine of the analytes exceeded the bias criterion within 72 h after initial measurement in either serum samples, plasma samples or both. Lithium-heparin plasma samples showed increasing values for phosphor, potassium and lactate dehydrogenase (LDH), which were only considered stable for respectively 24, 12 and 4 h, glucose was considered stable for 8 h. Electrolyte concentrations and LDH activity significantly increased in serum samples beyond 48 h. Bicarbonate should not be performed as add-on test at all.
Conclusions:
The presented data indicate that the conditions within an SRM have no clinical impact on sample stability and allow stable measurement of routine analytes within 72 h, comparable to manual storage facilities.
Related Concept Videos
Sample Handling
Samples should be transported carefully from collection points to the laboratory. They should be properly sealed and clearly labeled to prevent cross-contamination. To preserve the sample integrity, optimal temperature conditions during transport are essential. This could involve using...
Data Validation
Key parameters for method validation include:
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Drug Product Stability
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...