Agreement between oscillometric and direct blood pressure measurements in anesthetized captive chimpanzees (Pan
Christopher K Smith1, Anthony L Ashley1, Xiaojuan Zhu1
1From the Department of Small Animal Clinical Sciences (Smith, Cushing) and Office of Information and Technology (Zhu), College of Veterinary Medicine, University of Tennessee, Knoxville, TN 37996; and The Chattanooga Zoo, Chattanooga, TN 37404 (Ashley).
Objective:
To evaluate the level of agreement (LOA) between direct and oscillometric blood pressure (BP) measurements and the ability of oscillometric measurements to accurately detect hypotension in anesthetized chimpanzees (Pan troglodytes).
Animals:
8 captive, adult chimpanzees.
Procedures:
During prescheduled annual examinations, each chimpanzee underwent general anesthesia and patient monitoring for their examination, echocardiography for a concurrent study, and measurement of direct BP with the use of tibial artery catheterization and oscillometry with the use of a cuff placed around a brachium and a cuff placed around the second digit of the contralateral forelimb for the present study. Bland-Altman plots were generated to compare results for direct and oscillometric BP measurements. Mean bias and 95% LOAs were calculated for oscillometric measurements of systolic arterial pressure (SAP), diastolic arterial pressure (DAP), and mean arterial pressure (MAP) for each cuff site. Sensitivity and specificity in detecting hypotension were also determined for each cuff site.
Results:
There were 74 paired direct and brachial oscillometric measurements of each, SAP, MAP, and DAP and 66 paired direct and digit oscillometric measurements of each, SAP, MAP, and DAP. Only brachial oscillometric measurements of MAP had adequate sensitivity (78%) and specificity (95%) to accurately detect hypotension, and this technique also had the least mean bias (0.8 mm Hg; 95% LOA, -29 to 31 mm Hg).
Conclusions And Clinical Relevance:
Results indicated that brachial oscillometric measurement of MAP provided reasonable agreement with tibial arterial direct MAP measurement and performed well in diagnosing hypotension in anesthetized chimpanzees.
Related Concept Videos
Measurement of Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure