Exact analytical expressions for the band broadening in polydisperse 2-D multi-capillary columns with diffusional
Bram Huygens1, François Parmentier2, Gert Desmet1
1Department of Chemical Engineering, Vrije Universiteit Brussel, Pleinlaan 2, 1050 Brussels, Belgium.
Abstract:
Using a recent generalization of the Aris dispersion method, we have derived exact analytical expressions for the long-time limit dispersion in 2-D multi-capillary packings with diffusional bridging (binary channel system). Both the plug flow and the parabolic flow case are considered. The expressions are mathematically exact over the entire range of possible values of the degree of polydispersity (σ), the retention equilibrium constant K and the diffusion coefficients in the mobile and stationary zone. They are validated by comparing them to the dispersion data obtained by numerically solving the partial differential equation describing the general advection-diffusion mass balance. A correlation coefficient of R2 > 0.99995 is obtained. The form of the obtained analytical expression shows the dispersion arising from the polydispersity effect in multi-capillary systems can be split in two contributions, one related to the actual velocity difference between the adjacent channels and one related to the fact that the average of the intra-capillary C-term band broadening is larger in any σ≠0-case than in the σ=0-case. When leaving the small σ-assumption, a new factor ((1- σ2)/(1+3σ2)2) emerges, common to both the solution for the plug flow and the parabolic case. The analysis shows the parabolic flow not only leads to a larger intra-capillary dispersion (σ=0) than the plug flow case as known from literature but is also more sensitive to the polydispersity effect (up to some 10 to 20%).
Related Concept Videos
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Analyte Adsorption and Distribution