A Fluid-Management Drape for Hysteroscopy: Innovation for Improved Patient Safety and Surgical Care
Paul B Marshburn1, Brittany L Anderson-Montoya, Sally Baek
1Department of Obstetrics and Gynecology, the Division of Reproductive Endocrinology and Infertility, Quality and Patient Safety, the Carolinas Simulation Center, Innovation, the Center for Outcomes Research and Evaluation, and Strategy and Transformation, Atrium Health, Charlotte, North Carolina.
A novel total capture drape effectively collects fluid during hysteroscopy (uterine procedure), showing improved usability and accuracy for measuring patient fluid absorption compared to standard drapes.
Area of Science:
- Gynecologic surgery
- Medical device innovation
- Patient safety
Background:
- Accurate measurement of patient fluid absorption during hysteroscopy is crucial.
- Current methods using standard drapes may lead to fluid loss.
- A novel total capture drape was developed to address this challenge.
Purpose of the Study:
- To evaluate the effectiveness and usability of a novel total capture drape.
- To compare fluid-capture efficiency against a standard drape during hysteroscopy.
- To assess surgeon confidence and accuracy in fluid deficit measurement.
Main Methods:
- Simulation trials were conducted to assess fluid collection.
- An early-phase study randomized 68 women to compare total capture and standard drapes.
- Data collection included fluid deficit measurements, direct observation, and focus groups.
Main Results:
- Simulation showed complete fluid capture with the total capture drape versus progressive loss with the standard drape.
- The early-phase study found no statistical difference in fluid deficit but fewer cases of lost fluid with the total capture drape.
- Surgeons reported increased confidence in accuracy with the total capture drape after familiarization and training.
Conclusions:
- The total capture drape demonstrates favorable results, with comparable fluid collection to standard drapes.
- Repeated use and training enhance surgeon confidence in the total capture drape's accuracy.
- Design modifications are recommended to further improve usability and fluid-capture efficiency.