Changes in the Area Adjacent to the Internal Receiver-Stimulator of Cochlear Implant: A Retrospective Study
Isra Aljazeeri1, Tawfiq Khurayzi2, Yassin Abdelsamad3
1King Abdullah Ear Specialist Center, College of Medicine, King Saud University Medical City, King Saud University, Riyadh, Saudi Arabia | Aljabr Ophthalmology and Otolaryngology Hospital, Ministry of Health, Ahsa, Saudi Arabia.
Background:
In some patients with cochlear implants, bony resorption at the internal receiver- stimulator site can be observed on postoperative computed tomography. Therefore, it is essential to elucidate the effects of the internal receiver-stimulator on the scalp and bony bed over time.
Objective:
We aimed to evaluate how the internal receiver-stimulator of a cochlear implant device changed the thickness of the surrounding scalp and skull over time using computed tomography.
Methods:
This retrospective study evaluated patients who underwent cochlear implantation and received two computed tomography scans postoperatively for different indications at a tertiary referral centre. The main outcomes were scalp thickness and bony bed depth.
Results:
Fourteen ears were included in this study. There was very good inter-rater reliability among the two readers who evaluated the computed tomography of the included patients, with a Cronbach's alpha of 0.94. The mean scalp thickness over the internal receiver-stimulator was 6.02 (+/-2.4) mm in the first scan and decreased with no significant change to 5.62 mm (+/-1.64) in the second scan (p = 0.59, paired t-test). The mean depth of the bony bed increased significantly from 1.39 mm (+/-0.93) to 2.62 mm (+/-1.24) (p = 0.03).
Conclusion:
There was no change in the thickness of the scalp overlying the internal receiver-stimulator indicating that the scalp was more resistant than the bony skull to the tensile pressure exerted by the internal receiver-stimulator. In contrast, the bony bed depth of the internal receiver-stimulator increased over time. This can result in decreased internal receiver-stimulator protrusion and decreased risk of device displacement and migration.