Incidental operating room fire from a breathing circuit warmer system: a case report
Wangseok Do1, Dahyun Kang1, Purna Hong1
1Biomedical Research Institute, Pusan National University Hospital, Busan, 49241, Republic of Korea.
Background:
An airway-associated fire in an operating room can have devastating consequences for patients. Breathing circuit warmers (BCWs) are widely used to provide heated and humidified anesthetic gases and eventually prevent hypothermia during general anesthesia. Herein, we describe a case of a BCW-related airway fire.
Case Presentation:
In this case, an electrical short within a BCW wire caused a fire inside the circuit. Simultaneously, the fire was extinguished, ventilation was stopped, and the endotracheal tube was disconnected from the BCW. The patient was exposed to the fire for less than 10 s, resulting in burns to the trachea and bronchi. Immediately after airway burn, bronchoscopy showed no edema or narrowing except for soot in the trachea and both main bronchus. After the inhalation burn event, prophylactic antibiotics, bronchodilator, mucolytics nebulizer, and corticosteroid nebulizer were started. On bronchoscopy 3 days after the inhalation burn, mucosal erythematous edema was observed and the inflammatory reaction worsened. The inflammatory reaction showed aggravation for up to 2 weeks, and then gradually recovered, and the epithelium and mucous membrane of the upper respiratory tract returned to normal after 4 weeks. Eventually, the patient recovered without long-term complications and was successfully discharged.
Conclusions:
This is the first report of a fire caused by BCW. We wanted to share our experience of how we responded to an airway-related fire in an OR and treated the patient. It cannot be overemphasized that the electrical medical appliance associated with the airways are fatal to the patient in the event of a fire, so caution should always be exercised.
Related Concept Videos
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Acute Respiratory Failure-III
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Acute Respiratory Failure-IV