Related Experiment Video
Updated: Oct 14, 2025

Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Differences in the thermal properties and surface temperature of prehospital antihypothermia devices: an in vitro
Erez Dvir1,2, Danny Epstein3,4, Baruch Berzon5
1Department of Anaesthesiology, Rambam Health Care Campus, Haifa, Israel.
Background:
Preventing and treating hypothermia in prehospital settings is crucial. Several products have been developed to prevent heat loss and actively warm patients in prehospital settings. We compared the efficacy and the surface temperature of different antihypothermia products, using a fluid-based model at two ambient temperatures.
Methods:
We tested five active (Blizzard Heat with active pads, Ready-Heat, Ready-Heat-II, Hypothermia Prevention and Management Kit (HPMK), Bair Hugger) and five passive (Blizzard Heat, Heat Reflective Shell, sleeping bag, 'space blanket', wool blanket) antihypothermia products. A torso model consisting of four 8 L bags of fluid preheated to 36°C±0.5°C (97±0.5°F) was used to compare the devices' performances at 20°C (68°F) and 8°C (46°F). Inner and surface temperatures were recorded for up to 480 min.
Results:
We found significant differences in heat loss in fluid bags among the tested devices at both temperatures (p<0.001). At 20°C, only HPMK and Ready-Heat-II increased the inner temperature for 480 min while Blizzard Heat with active pads prevented heat loss. Ready-Heat prevented heat loss for 90 min. All the other devices did not prevent heat loss beyond 30 min. At 8°C, none of the products heated the model. Bair Hugger, HPMK, Ready-Heat II and sleeping bag prevented heat loss for 30 min. At 60, 90 and 120 min HPMK, Ready-Heat II and Bair Hugger were the most effective. Over 480 min, Bair Hugger was most effective, with a heat loss of 2.3°C±0.4°C. The surface temperature exceeded 44°C (111°F) for all the exothermic warming devices when used for a prolonged period of time.
Conclusion:
At 20°C, HPMK and Ready-Heat-II increased fluid temperature in the model, while the other devices decreased heat loss. At 8°C, none of the tested devices increased the temperature. However, active heating devices prevented heat loss slightly better than passive methods. A protective insulation layer should be used with all active heating blankets.
More Related Videos
05:00Short-Duration Hypothermia Induction in Rats using Models for Studies examining Clinical Relevance and Mechanisms
Published on: March 3, 2021
08:22In vitro Assessment of Myocardial Protection following Hypothermia-Preconditioning in a Human Cardiac Myocytes Model
Published on: October 27, 2020
Related Concept Videos
Decreased Body Temperature
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...