Comparison of culture, confocal microscopy and PCR in routine hospital use for microbial keratitis diagnosis
Jeremy J Hoffman1,2,3, John K G Dart4,5, Surjo K De4,6
1Moorfields Eye Hospital NHS Foundation Trust, London, UK. j.hoffman@ucl.ac.uk.
Aims:
To evaluate the sensitivity and specificity of polymerase chain reaction (PCR), in vivo confocal microscopy (IVCM) and culture for microbial keratitis (MK) diagnosis.
Methods:
Retrospective review of PCR, IVCM and culture results for MK diagnosis at Moorfields Eye Hospital between August 2013 and December 2014.
Results:
PCR results were available for 259 MK patients with concurrent culture for 203/259 and IVCM for 149/259. Sensitivities and specificities with 95% confidence intervals [95% CI] were calculated for Acanthamoeba keratitis (AK) and fungal keratitis (FK), by comparison with culture, for both IVCM and PCR. For AK, FK and bacterial keratitis (BK) sensitivities were calculated, for each diagnostic method, by comparison with a composite reference standard (a positive result for one or more of culture, PCR or IVCM having a specificity of 100% by definition). For the latter, sensitivities with [95% CI] were: for AK, IVCM 77.1% [62.7-88.0%], PCR 63.3% [48.3-76.6%], culture 35.6 [21.9-51.2]; for FK, IVCM 81.8% [48.2-97.7%], PCR 30.8% [9.09-61.4%], culture 41.7% [15.2-72.3%]; for BK, PCR 25.0% [14.7-37.9%], culture 95.6% [87.6-99.1%].
Conclusion:
IVCM was the most sensitive technique for AK and FK diagnosis but culture remains our gold standard for BK. These findings reflect results to be expected from service providers to UK ophthalmology units and demonstrates the need at our centre for ongoing diagnostic result audit leading to the potential to improve PCR diagnosis. Both FK and AK are now common in the UK; ophthalmology units need to have all these techniques available to optimise their MK management.
Related Concept Videos
Confocal Fluorescence Microscopy
Three-Dimensional Microscopy in Microbiology
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...