Successful application of closed loop stimulation pacemakers with remote monitoring in 3 miniature donkeys with
Lisa De Lange1, Glenn Van Steenkiste1, Ingrid Vernemmen1
1Equine Cardioteam, Department of Large Animal Internal Medicine, Faculty of Veterinary Medicine, Merelbeke, Belgium.
Abstract:
Rate-adaptive single chamber pacemakers with accelerometer, closed loop stimulation (CLS), and remote monitoring functionality (Eluna 8 SR-T, Biotronik, SE & Co, Germany) were implanted in 3 miniature donkeys with third-degree atrioventricular block and syncope. After recovery, different pacemaker programming modes were tested at rest, during stress without physical exercise and during physical exercise. Pacing rates were compared to actual atrial rates and showed that CLS functionality allowed physiological heart rate adaptation. A transmitter installed in the stable provided wireless connection of the pacemaker to the internet. Home monitoring was activated which performed daily wireless transmission of pacemaker functional measurements to an online server allowing diagnosis of pathological arrhythmias and pacemaker malfunction from a distance. Closed loop stimulation and remote monitoring functionality resulted in nearly physiological rate adaptation and allowed remote "from-the-stable" patient follow-up.