Unusual MR metallic artifact due to steel threads
Journal of Computer Assisted Tomography
|July 1, 1987
Abstract:
Metallic objects are a well known source of artifact in magnetic resonance imaging. We report a case of artifacts created by metallic fibers woven into surgical scrub pants worn by a patient during the examination.
Insights
Metallic fibers in surgical scrub pants caused significant artifact in magnetic resonance imaging (MRI). This case highlights the importance of patient clothing in MRI safety and image quality.
Area of Science:
- Medical Imaging
- Radiology
- Materials Science
Background:
- Metallic objects are known sources of artifact in Magnetic Resonance Imaging (MRI).
- Artifacts can degrade image quality and potentially lead to misdiagnosis.
- Standard protocols often involve screening for metallic implants and accessories.
Observation:
- A patient undergoing MRI examination presented with unexpected and severe image artifacts.
- The artifacts were traced to metallic fibers incorporated into the fabric of surgical scrub pants worn by the patient.
- The metallic fibers, though not overtly visible, interacted with the MRI's magnetic field.
Findings:
- The metallic fibers in the surgical scrub pants generated significant signal distortions and geometric inaccuracies in the MRI scans.
- The pattern of artifacts was consistent with those produced by metallic foreign bodies.
- The presence of these fibers, even when not in direct contact with the skin, was sufficient to cause substantial image degradation.
Implications:
- Patient clothing should be considered a potential source of MRI artifacts, in addition to implants and accessories.
- Radiology departments may need to update screening protocols to include inquiries about clothing materials.
- Awareness of this artifact source can improve diagnostic accuracy and patient safety in MRI examinations.