Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
Limit of detection comparison on urine gross alpha/beta, H-3, and P-32 analysis between different liquid
Olga Piraner1, Robert L Jones1
1Centers for Disease Control and Prevention, National Center for Environmental Health, Division of Laboratory Sciences, Inorganic and Radiation Analytical Toxicology Branch 4770 Buford Hwy, MS S110-5, Atlanta, GA 30341-3717.
Abstract:
As part of the Centers for Disease Control and Prevention's post-radiological/nuclear incident response mission, we developed rapid bioassay analytical methods to assess possible human exposure to radionuclides and internal contamination. Liquid scintillation counting (LSC) is a valuable analytical tool for the rapid detection and quantification of gross alpha/beta-emitting radionuclides in urine samples. A key characteristic of this type of bioassay method is its detection sensitivity for the priority threat radionuclides. We evaluated the limit of detection of selected LSC instruments to determine which instrument can be used when low-dose measurement is important.