When synchronized is not synchronous: an experimental benchmarking study on the efficiency of SIMV in very-low-birth
Lorenz Auer-Hackenberg1, Clemens Haselmann2, Johannes Brandner2
1Division of Neonatology, Department of Pediatrics and Adolescent Medicine, Paracelsus Medical University, Salzburg, Austria - l.auer-hackenberg@salk.at.
Background:
Synchronized ventilation promotes a patient's ability to breathe spontaneously by providing intermittent, mechanical-controlled respiration that is synchronized with the patient's own efforts. In synchronized-intermittent-mandatory-ventilation (SIMV), assisted ventilation is regulated by frequency settings which dictate the interval at which the ventilator becomes sensitive to respiratory efforts and responds with an assisted breath. SIMV has become one of the most widely used modes of ventilation in neonates. Using a neonatal-active-lung-model (NALM), this in-vitro benchmark study investigated how well synchronization works in SIMV with several ventilators.
Methods:
The competence of eight ventilators was tested using a NALM simulator representing a preterm infant weighing approximately 1500 grams. Two conditions were explored: first, the ventilators were set to a constant ventilation rate, while the NALM was adjusted to frequencies equal to and below this ventilation rate. The second condition varied the ventilators' rates while the NALM frequency was held constant. Correctly triggered breaths were counted and displayed as a percentage (%) of the total potential triggerable breaths.
Results:
Performance among devices significantly differed, ranging from a low 38.9% competency to a max of 71.7% under the first condition, and 70.7% to 100% under the second condition.
Conclusions:
At high SIMV frequencies, synchronization between the patient and ventilator becomes increasingly limited. Despite their identical ventilator functions, SIMV algorithms of the various manufacturers and models tested, deliver ventilation rates with significantly different degrees of synchronization; not only in comparison to each other, but also in their own ability to continuously and effectively synchronize breaths under variable conditions, typical of preterm lungs.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...