Multicenter Clinical Evaluation of ETEST Plazomicin (PLZ) for Susceptibility Testing of Enterobacterales

Laurine S Blanchard1, Alex Van Belkum2, Dominique Dechaume2

  • 1bioMérieux SA, Marcy l'Etoile, France.

Insights

The ETEST Plazomicin test accurately determines susceptibility for Enterobacterales, offering a reliable alternative to traditional methods for this novel antibiotic.

Area of Science:

  • Clinical Microbiology
  • Infectious Diseases
  • Pharmacology

Background:

  • Plazomicin (PLZ), marketed as ZEMDRI, is a novel aminoglycoside antibiotic approved for complicated urinary tract infections.
  • Antimicrobial susceptibility testing (AST) is crucial for guiding treatment, with broth micro-dilution (BMD) being a standard but labor-intensive method.
  • Gradient diffusion methods, like ETEST®, offer a potentially more convenient alternative for AST.

Purpose of the Study:

  • To evaluate the performance of the novel ETEST Plazomicin (PLZ) assay.
  • To compare ETEST PLZ results against the gold standard broth micro-dilution (BMD) method for Enterobacterales.
  • To assess ETEST PLZ accuracy according to FDA and ISO guidelines using established breakpoints.

Main Methods:

  • A multi-center study evaluated the ETEST PLZ against BMD for AST.
  • 598 clinical isolates of Enterobacterales were tested, including 53 resistant to PLZ by BMD.
  • Performance metrics included essential agreement (EA), category agreement (CA), and error rates (VME, ME, mE).

Main Results:

  • ETEST PLZ demonstrated high overall essential agreement (99.0%) and category agreement (92.8%) with BMD.
  • The test showed a low rate of very major errors (1.9%) and no major errors (0%).
  • Individual species exhibited essential agreement rates of 90% or higher, with a single very major error in Serratia marcescens.

Conclusions:

  • The ETEST Plazomicin assay is an accurate tool for performing antimicrobial susceptibility testing of Enterobacterales.
  • ETEST PLZ provides a reliable and potentially more efficient alternative to BMD for guiding plazomicin therapy.
  • The findings support the clinical utility of ETEST PLZ for routine AST in healthcare settings.