Heart assist devices and their impact on arterial Doppler: When the waveform appears flat or serrated
Yi Wang1,2
1Alfred Imaging Group 144 Great North Road Five Dock New South Wales 2046 Australia.
Objective:
The purpose of this article was to review the principal mechanisms of three types of heart assist devices, to explore their haemodynamic impacts on spectral Doppler and to discuss the associated challenges in diagnostic ultrasound.
Conclusion:
Heart assist devices are used in patients with advanced heart failure as bridge to transplantation or destination therapy. The arterial Doppler waveform features can be altered after the implantation of the devices. A continuous-flow device converts the arterial waveform into a venous-like waveform with a mean velocity that can be used in stenosis evaluation. When the pulsatile device produces extra pulses between the natural heartbeats, together they produce serrated arterial waveforms. When the pulsatile device serves as the only pumping source, the arterial waveforms are similar to the natural ones. Colour Doppler features are maintained despite all the spectral Doppler changes. These devices can affect both ultrasound scanning and interpretation. Therefore, it is essential to understand the principal mechanisms of these devices and their haemodynamic impacts so that an accurate ultrasound diagnosis can be made.
More Related Videos
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Cardiovascular System Abnormal Findings I: Inspection and Palpation
Abnormal findings observed during an inspection
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Pulse
The pulse serves as a clinical...