Background:
Biological safety of ultrasound is a complex and nuanced subject that is poorly understood by ultrasound users. Little is known about the acoustic output and thermal index levels during the routine use of modern ultrasound machines in prenatal scanning.
Methodology:
This study was a retrospective review of thermal index (TI) values encountered during 300 prenatal ultrasound examinations (100 in each trimester) performed on any one of 13 Philips Epiq 7 or Epiq 5 systems, representing approximately 106.5 h of real-time scanning. The TI levels were compared to three international guidelines on the biological safety of ultrasound.
Results:
The routine use of current Philips systems was associated with low TI levels. Of the 300 examinations reviewed, virtually all were compliant with the BMUS and Nelson safety guideline. Whether the examination was compliant with the WFUMB guideline is open to interpretation. The highest level of TI encountered was 1.1. In no instance did the TI level incur into the 'not recommended' range or into a range where specific user action was required to reduce the TI within 1 min. The most frequent action associated with TI > 0.7 was the use of M-mode to document the fetal heart rate. In the four instances where TI peaked at 1.1, 3 were associated with the use of M-mode and one with B-mode. Spectral Doppler was not implicated in high TI levels. These results are surprising and open up a range of opportunities for future study.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Imaging Studies II: Ultrasonography
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: