Stacey J L Sullivan1, Jean E Rinaldi1, Prasanna Hariharan2
1Division of Applied Mechanics, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, United States Food and Drug Administration, 10903 New Hampshire Ave, Silver Spring, MD, 20993, USA.
Abstract:
Non-contact infrared thermometers (NCITs) are being widely used during the COVID-19 pandemic as a temperature-measurement tool for screening and isolating patients in healthcare settings, travelers at ports of entry, and the general public. To understand the accuracy of NCITs, a clinical study was conducted with 1113 adult subjects using six different commercially available NCIT models. A total of 60 NCITs were tested with 10 units for each model. The NCIT-measured temperature was compared with the oral temperature obtained using a reference oral thermometer. The mean difference between the reference thermometer and NCIT measurement (clinical bias) was different for each NCIT model. The clinical bias ranged from just under - 0.9 °C (under-reporting) to just over 0.2 °C (over-reporting). The individual differences ranged from - 3 to + 2 °C in extreme cases, with the majority of the differences between - 2 and + 1 °C. Depending upon the NCIT model, 48% to 88% of the individual temperature measurements were outside the labeled accuracy stated by the manufacturers. The sensitivity of the NCIT models for detecting subject's temperature above 38 °C ranged from 0 to 0.69. Overall, our results indicate that some NCIT devices may not be consistently accurate enough to determine if subject's temperature exceeds a specific threshold of 38 °C. Model-to-model variability and individual model accuracy in the displayed temperature were found to be outside of acceptable limits. Accuracy and credibility of the NCITs should be thoroughly evaluated before using them as an effective screening tool.
More Related Videos
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
08:16Infrared Thermography for the Detection of Changes in Brown Adipose Tissue Activity
Published on: September 28, 2022
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...