Combined Compressed Sensing and SENSE to Enhance Radiation Therapy Magnetic Resonance Imaging Simulation
Victoria Y Yu1, Kristen Zakian1, Neelam Tyagi1
1Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, New York.
Purpose:
To assess the effect of a combination of compressed sensing and SENSitivity Encoding (SENSE) acceleration techniques on radiation therapy magnetic resonance imaging (MRI) simulation workflows.
Methods And Materials:
Thirty-seven acquisitions were performed with both SENSE-only (SENSE) and combined compressed sensing and SENSE (CS) techniques in 24 patients receiving radiation therapy MRI simulation for a wide range of disease sites. The anatomic field of view prescription and image resolution were identical for both SENSE and CS acquisitions to ensure fair comparison. The acquisition time of all images was recorded to assess time savings. For each image pair, image quality, and ability to contour were assessed by 2 radiation oncologists. Aside from direct image pair comparisons, the feasibility of using CS to improve MRI simulation protocols by increasing image resolution, field of view, and reducing motion artifacts was also evaluated.
Results:
CS resulted in an average reduction of 27% in scan time with negligible changes in image quality and the ability to contour structures for RT treatment planning compared with SENSE. Physician scoring of image quality and ability to contour shows that while SENSE still has slightly better image quality compared with CS, this observed difference in image quality did not affect the ability to contour. In addition, the higher acceleration capability of CS enabled use of superior-inferior direction phase encoding in a sagittal 3-dimensional T2-weighted scan for substantially improved visibility of the prostatic urethra, which eliminated the need for a Foley catheter in most patients.
Conclusions:
The combination of compressed sensing and parallel imaging resulted in marked improvements in the MRI Simulation workflow. The scan time was reduced without significantly affecting image quality in the context of ability to contour. The acceleration capabilities allowed for increased image resolution under similar scanning times as well as significantly improved urethra visualization in prostate simulations.
More Related Videos
Related Concept Videos
Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...