Development of a point-of-care test for the detection of MDMA in latent fingerprints using surface plasmon resonance

Caroline Pollard1, Mark Hudson2, James M McDonnell3

  • 1King's Forensics, Department of Analytical, Environmental & Forensic Sciences, Franklin-Wilkins Building, King's College London, 150 Stamford St, London, United Kingdom, SE1 9NH, UK.

Drug Testing and Analysis
|November 12, 2021
PubMed

Insights

A new point-of-care test (POCT) detects 3,4-methylenedioxymethamphetamine (MDMA) in latent fingerprints. This sensitive test uses lateral flow immunoassay technology for rapid screening without sample preparation.

Area of Science:

  • Forensic Science
  • Analytical Chemistry
  • Biotechnology

Background:

  • No specific point-of-care test (POCT) exists for detecting 3,4-methylenedioxymethamphetamine (MDMA) in latent fingerprints (LFPs).
  • Existing POCTs for MDMA in sweat often detect it as a cross-reactant, reducing sensitivity.

Purpose of the Study:

  • To design a sensitive POCT for MDMA detection in LFPs.
  • To utilize surface plasmon resonance (SPR) and lateral flow immunoassay (LFA) technologies.

Main Methods:

  • Identified a high-affinity antibody pair using SPR for LFA suitability.
  • Optimized antibody and antigen concentrations for clear positive/negative signal differentiation.
  • Tested LFA performance with dose-response curves for MDMA and structurally similar drugs.

Main Results:

  • Developed an LFA with a cut-off of 60 pg/10 μl, achieving 1.00 sensitivity, 0.95 specificity, and 0.98 accuracy.
  • Determined cross-reactivity levels, with PMMA showing the highest at 250%.
  • Successfully created a sensitive LFP screening tool requiring no sample preparation.

Conclusions:

  • A novel, sensitive POCT for MDMA in LFPs was successfully designed and validated.
  • The developed LFA offers rapid, specific detection without the need for sample preparation.
  • This technology has potential applications in forensic and law enforcement settings.