Clinical Validation of the Accuracy of Absolute Myocardial Blood Flow Quantification with Dual-Source CT Using
Masafumi Takafuji1, Kakuya Kitagawa1, Masaki Ishida1
1Department of Radiology (M.T., K.K., M.I., Y.I., S. Nakamura, H.S.) and Department of Cardiology and Nephrology (S. Nakamori, T.K., K.D.), Mie University Hospital, 2-174 Edobashi, Tsu, Mie 514-8507, Japan.
Purpose:
To determine the fitting equation that can correct for the underestimation of myocardial blood flow (MBF) measurement by using dynamic CT perfusion (CTP) with dual-source CT (MBFCT), using MBF with oxygen 15-labeled water (15O-water) PET (MBFPET) as a reference, and to determine the accuracy of corrected MBFCT (MBFCT-corrected) compared with MBFPET in a separate set of participants.
Materials And Methods:
In this prospective study (reference no. 2466), 34 participants (mean age, 70 years ± 8 [standard deviation]; 27 men) known or suspected to have coronary artery disease underwent dynamic stress CTP and stress 15O-water PET between January 2014 and December 2018. The participants were randomly assigned to either a pilot group (n = 17), to determine the fitting equation on the basis of the generalized Renkin-Crone model that can explain the relation between MBFCT and MBFPET, or to a validation group (n = 17), to validate MBFCT-corrected compared with MBFPET. The agreement between MBFCT-corrected and MBFPET was evaluated by intraclass correlation and Bland-Altman analysis.
Results:
In the pilot group, MBFCT was lower than MBFPET (1.24 mL/min/g ± 0.28 vs 2.51 mL/min/g ± 0.89, P < .001) at the segment level. The relationship between MBFCT and MBFCT-corrected was represented as MBFCT = MBFCT-corrected × {1-exp[-(0.11 × MBFCT-corrected + 1.54)/MBFCT-corrected]}. In the validation group, MBFCT-corrected was 2.66 mL/min/g ± 1.93, and MBFPET was 2.68 mL/min/g ± 1.87 at the vessel level. MBFCT-corrected showed an excellent agreement with MBFPET (intraclass correlation coefficient = 0.93 [95% CI: 0.87, 0.96]). The measurement bias of MBFCT-corrected and MBFPET was -0.02 mL/min/g ± 0.74.
Conclusion:
Underestimation of MBF by CT was successfully corrected with a correction method that was based on contrast kinetics in the myocardium.Keywords: CT, CT-Perfusion, PET, Cardiac, Heart Supplemental material is available for this article. © RSNA, 2021.