An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Long-term evaluation of factors affecting removal torque of microimplants
1Department of Orthodontics, School of Dentistry, Kyungpook National University, 2175, Dalgubeoldae-ro, Jung-Gu, Daegu, 41940, Korea.
Background:
The current study aimed to evaluate factors affecting the long-term stability of microimplants using removal torque and the correlation between removal torque and clinical variables.
Materials And Methods:
This research evaluated 703 microimplants placed in 354 patients (mean age: 30.4 ± 12.1 years). The removal torque was evaluated according to various clinical variables including sex, age, placement site, microimplant size, and placement method (self-drilling versus pre-drilling). Pearson correlation and stepwise multiple linear regression analyses were performed to investigate different variables and their association with removal torque.
Results:
The mean removal torque was significantly higher in the mandible (4.46 N cm) than in the maxilla (3.73 N cm). The values in the posterior teeth/retromolar areas were significantly higher than those in the anterior teeth area. There were no significant difference in terms of sex. Teenagers had a lower removal torque than older adults in the mandible, but not in the maxilla. Microimplants with a greater length and diameter, except for those with a greater diameter in the maxilla, was associated with a higher removal torque. Regardless of placement torque, the removal torque convergently reached approximately 4 N cm in both placement methods. The removal torque was significantly correlated with screw length in the self-drilling group and with diameter in the pre-drilling group.
Conclusions:
Removal torque was related with placement site, age, placement method, and length and diameter of microimplants.
More Related Videos
06:36Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
11:38Environmentally-controlled Microtensile Testing of Mechanically-adaptive Polymer Nanocomposites for ex vivo Characterization
Published on: August 20, 2013
Related Concept Videos
Residual Stresses in Circular Shafts
Screw: Problem Solving
To evaluate the...
Circular Shafts - Elastoplastic Materials
As torque on the...