Particle therapy using protons or carbon ions for cancer patients with cardiac implantable electronic devices (CIED):
Takayuki Hashimoto1, Yusuke Demizu2, Haruko Numajiri3
1Department of Radiation Medical Science and Engineering, Faculty of Medicine, Hokkaido University, North 15 West 7, Kita-ku, Sapporo, Hokkaido, 060-8638, Japan. thashimoto@med.hokudai.ac.jp.
Purpose:
To evaluate the outcomes of particle therapy in cancer patients with cardiac implantable electronic devices (CIEDs).
Materials And Methods:
From April 2001 to March 2013, 19,585 patients were treated with proton beam therapy (PBT) or carbon ion therapy (CIT) at 8 institutions. Of these, 69 patients (0.4%, PBT 46, CIT 22, and PBT + CIT 1) with CIEDs (64 pacemakers, 4 implantable cardioverter defibrillators, and 1 with a cardiac resynchronization therapy defibrillator) were retrospectively reviewed. All the patients with CIEDs in this study were treated with the passive scattering type of particle beam therapy.
Results:
Six (13%) of the 47 PBT patients, and none of the 23 CIT patients experienced CIED malfunctions (p = 0.105). Electrical resets (7) and over-sensing (3) occurred transiently in 6 patients. The distance between the edge of the irradiation field and the CIED was not associated with the incidence of malfunctions in 20 patients with lung cancer. A larger field size had a higher event rate but the test to evaluate trends as not statistically significant (p = 0.196).
Conclusion:
Differences in the frequency of occurrence of device malfunctions for patients treated with PBT and patients treated with CIT did not reach statistical significance. The present study can be regarded as a benchmark study about the incidence of malfunctioning of CIED in passive scattering particle beam therapy and can be used as a reference for active scanning particle beam therapy.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Cardiomyopathy V: Interprofessional Care