Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Digital impressions' accuracy through "cut-out-rescan" and "data exchange by over scanning" techniques in complete
Leandro Passos1,2, Sergio Meiga3, Vinicius Brigagão4
1Department of Prosthetic Dentistry, Federal Fluminense University, Health Institute of Nova Friburgo, School of Dentistry, Rio de Janeiro, Brazil.
Purpose:
This technical procedure describes the accuracy of "cut-out-rescan" and "data exchange by over scanning" on cast areas using computer-aided design and computer-aided manufacturing software and two intraoral scanners (IOSs).
Methods:
A customized cast was used as a reference standard and scanned using an ATOS Triple Scan digitizer. Two IOS setups were used in three scanning groups, namely, the control, cut-out-rescan, and data exchange by over scanning groups. Sixty digital files (n = 10 per group) were obtained and converted into the standard tessellation language format. A metrology program was used to evaluate the difference in accuracy between the control group and the other groups using Welch's unequal variances t-test.
Conclusion:
For trueness and precision of complete arch scans, the present technique provides statistical evidence that using data exchange by over scanning and a Primescan IOS results in higher accuracy scans compared to using cut-out-rescan and a Primescan IOS. The cut-out-rescan procedure provides scans with higher accuracy when using an Omnicam IOS. This technique is simple and saves time in workflows using a Primescan IOS.