Numerical analysis of hemodynamic effect under different enhanced external counterpulsation (EECP) frequency for
Siwei Ye1, Ming Yang1, Yuanfei Zhu1
1Department of Instrument Science and Engineering, Shanghai Jiao Tong University, Shanghai, China.
Abstract:
Based on the changes in phase characteristics of blood flow and pressure, enhanced external counterpulsation (EECP) reduces cardiac load and improves cerebral perfusion in patients with cerebrovascular diseases. However, increased cerebral blood flow (CBF) is associated with the rise in blood pressure and its complications. Increased EECP frequency is a valuable solution when combined with the electrical equivalent impedance characteristics of the lumped parameter model (LPM) of the human blood circulation system. Herein, to investigate the effect of different EECP frequencies on CBF perfusion, an LPM was established with cardiopulmonary circulation and eight systemic blood flow units with cerebral autoregulation module of ischemic stroke patients. Then, using differential equations, we analyzed those parameters through hemodynamic simulations in four EECP modes. With related influencing parameters remaining constant, we adjusted the pressure frequency of EECP and found that when compared to the traditional sequential EECP mode, the relative increase rate of CBF was 16.68%, 18.95%, and 21.21% from 1 to 3 Hz, respectively. This study validates the effect of improving blood prefusion with increasing EECP frequency through numerical analysis.
More Related Videos
11:26Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression
Published on: December 10, 2014
10:41Analysis of Cerebral Vasospasm in a Murine Model of Subarachnoid Hemorrhage with High Frequency Transcranial Duplex Ultrasound
Published on: June 3, 2021
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...
