Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Continuous measurement of arterial oxygenation in mechanically ventilated horses
Emma V Hummer1, Joao H N Soares2, Douglas C Crockett3
1William Pritchard Veterinary Medical Teaching Hospital, University of California Davis, Davis, California, USA.
Background:
The possibility of accurately and continuously measuring arterial oxygen partial pressure (PaO2 ) in horses may facilitate the management of hypoxaemia during general anaesthesia.
Objectives:
The aim of this study was to evaluate the ability of a novel fibreoptic sensor to measure PaO2 (PaO2Sensor ) continuously and in real time in horses undergoing ventilatory manoeuvres during general anaesthesia.
Study Design:
In vivo experimental study.
Methods:
Six adult healthy horses were anaesthetised and mechanically ventilated in dorsal recumbency. A fibreoptic sensor was placed in one of the facial arteries through a catheter to continuously measure and record PaO2Sensor . After an alveolar recruitment manoeuvre, a decremental positive end-expiratory pressure (PEEP) titration using 20-minute steps of 5 cm H2 O from 20 to 0 cm H2 O was performed. An arterial blood sample was collected at 15 minutes of ventilation at each PEEP level for PaO2 measurement using an automated blood gas machine (PaO2Ref ). The agreement between PaO2Sensor and PaO2Ref was assessed by Pearson's correlation, Bland-Altman plot and four-quadrant plot analysis. In the last minute of ventilation at each PEEP level, a slow tidal inflation/deflation manoeuvre was performed.
Results:
The mean relative bias between PaO2Sensor and PaO2Ref was 4% with limits of agreement between -17% and 29%. The correlation coefficient between PaO2Sensor and PaO2Ref was 0.98 (P < .001). The PaO2Sensor and PaO2Ref concordance rate for changes was 95%. Measurements of PaO2Sensor during the slow inflation/deflation manoeuvre at PEEP 15 and 10 cm H2 O were not possible because of significant noise on the PaO2 signal generated by a small blood clot.
Main Limitations:
Small sample size.
Conclusion:
The tested fibreoptic probe was able to accurately and continuously measure PaO2Sensor in anaesthetised horses undergoing ventilatory manoeuvres. A heparinised system in the catheter used by the fibreoptic sensor should be used to avoid blood clots and artefacts in the PaO2 measurements.
More Related Videos
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...