The misunderstood meander: Redesigning MRI meander-line surface coils to reduce noise, increase uniformity, and
Gary Zabow1, Stephen Dodd2, Alan Koretsky2
1Applied Physics Division, National Institute of Standards and Technology, Boulder, CO 80305, USA.
Abstract:
Meander-line, or zig-zag, MRI surface coils theoretically promise spatially uniform fields with optimal field localization close to the coil. In reality, they suffer poorer than expected field localizations and acquired images are often highly inhomogeneous, plagued by repeating stripe-like signal-loss artifacts. We show that both these detrimental effects arise from coil design based on the same invalid approximation in the underlying theory. Here, the conventional approximation is corrected, yielding a modified coil design that validates the new theory by rectifying the above problems. Specifically, an easily implementable coil correction, which amounts to the addition of a single extra turn of wire, is introduced and shown to increase signal uniformity by an order of magnitude, eliminate image artifacts, and reduce unwanted signal interference from deeper within the sample by tightening the coil field localization to close to the coil, as intended for zig-zag designs. With independent optimization of coil size and imaging depth possible, such corrected meander-lines surface coils may be well suited for large area, near-surface imaging and spectroscopy applications.
Related Concept Videos
Magnetic Resonance Imaging
NMR Spectrometers: Resolution and Error Correction
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...