Related Experiment Video
Updated: Oct 12, 2025

Endotracheal Intubation Using a Flexible Intubation Endoscope as a Standardized Model for Safe Airway Management in Swine
Published on: August 25, 2022
Intubation using VieScope vs. Video laryngoscopy in full personal protective equipment - a randomized, controlled
Hannes Ecker1,2, Simone Kolvenbach1,2, Holger Herff1,2
1Faculty of Medicine, University of Cologne, Kerpener Str. 62, 50937, Cologne, Germany.
Background:
VieScope is a new type of laryngoscope, with a straight, transparent and illuminated blade, allowing for direct line of sight towards the larynx. In addition, VieScope is disposed of after single patient use, which can avoid cross-contaminations of contagious material. This has gained importance especially when treating patients with highly contagious infectious diseases, such as during the SARS-CoV2 pandemic. In this context, VieScope has not been evaluated yet in a clinical study.
Material And Methods:
This study compared intubation with VieScope to video-laryngoscopy (GlideScope) in normal and difficult airway in a standardized airway manikin in a randomized controlled simulation trial. Thirty-five medical specialists were asked to perform endotracheal intubation in full personal protective equipment (PPE). Primary endpoint was correct tube position. First-pass rate (i.e., success rate at the first attempt), time until intubation and time until first correct ventilation were registered as secondary endpoints.
Results:
For correct tracheal tube placement, there was no significant difference between VieScope and GlideScope in normal and difficult airway conditions. VieScope had over 91% fist-pass success rate in normal airway setting. VieScope had a comparable success rate to GlideScope in difficult airway, but had a significantly longer time until intubation and time until ventilation.
Conclusion:
VieScope and GlideScope had high success rates in normal as well as in difficult airway. There was no unrecognized esophageal intubation in either group. Overall time for intubation was longer in the VieScope group, though in an acceptable range given in literature. Results from this simulation study suggest that VieScope may be an acceptable alternative for tracheal intubation in full PPE.
Trial Registration:
The study was registered at the German Clinical Trials Register www.drks.de (Registration date: 09/11/2020; TrialID: DRKS00023406 ).
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Tracheostomy Suctioning II: Procedure
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Endoscopic Procedures III: Video Capsule Endoscopy