Image quality in liver CT: low-dose deep learning vs standard-dose model-based iterative reconstructions
Sungeun Park1,2, Jeong Hee Yoon1,3,4, Ijin Joo1,3
1Department of Radiology, Seoul National University Hospital, 101 Daehak-ro, Jongno-gu, Seoul, 03080, Republic of Korea.
Objectives:
To compare the overall image quality and detectability of significant (malignant and pre-malignant) liver lesions of low-dose liver CT (LDCT, 33.3% dose) using deep learning denoising (DLD) to standard-dose CT (SDCT, 100% dose) using model-based iterative reconstruction (MBIR).
Methods:
In this retrospective study, CT images of 80 patients with hepatic focal lesions were included. For noninferiority analysis of overall image quality, a margin of - 0.5 points (scored in a 5-point scale) for the difference between scan protocols was pre-defined. Other quantitative or qualitative image quality assessments were performed. Additionally, detectability of significant liver lesions was compared, with 64 pairs of CT, using the jackknife alternative free-response ROC analysis, with noninferior margin defined by the lower limit of 95% confidence interval (CI) of the difference of figure-of-merit less than - 0.1.
Results:
The mean overall image quality scores with LDCT and SDCT were 3.77 ± 0.38 and 3.94 ± 0.34, respectively, demonstrating a difference of - 0.17 (95% CI: - 0.21 to - 0.12), which did not cross the predefined noninferiority margin of - 0.5. Furthermore, LDCT showed significantly superior quantitative results of liver lesion contrast to noise ratio (p < 0.05). However, although LDCT scored higher than the average score in qualitative image quality assessments, they were significantly lower than those of SDCT (p < 0.05). Figure-of-merit for lesion detection was 0.859 for LDCT and 0.878 for SDCT, showing noninferiority (difference: - 0.019, 95% CI: - 0.058 to 0.021).
Conclusion:
LDCT using DLD with 67% radiation dose reduction showed non-inferior overall image quality and lesion detectability, compared to SDCT.
Key Points:
• Low-dose liver CT using deep learning denoising (DLD), at 67% dose reduction, provided non-inferior overall image quality compared to standard-dose CT using model-based iterative reconstruction (MBIR). • Low-dose CT using DLD showed significantly less noise and higher CNR lesion to liver than standard-dose CT using MBIR and demonstrated at least average image quality score among all readers, albeit with lower scores than standard-dose CT using MBIR. • Low-dose liver CT showed noninferior detectability for malignant and pre-malignant liver lesions, compared to standard-dose CT.
More Related Videos
06:09Author Spotlight: Advancing Hepatic Fibrosis Diagnosis Using Magnetic Resonance Elastography and AI
Published on: July 21, 2023
04:09Predicting Treatment Response to Image-Guided Therapies Using Machine Learning: An Example for Trans-Arterial Treatment of Hepatocellular Carcinoma
Published on: October 10, 2018
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT